Pressure sensitivity (ratio of dynamic output voltage to dynamic pressure), typically expressed as a logarithmic level quantity relative to 1 V/Pa.
The measurement capability (primary calibration via reciprocity method ISO 61094-2) is traceable to fundamental physical quantities:
AC voltage
Length (Dimensions of microphone and couplers)
Capacitance (reference capacitance for transfer impedance)
Mass (air density)
Temperature
Frequency
Time
Pressure (static pressure)
Nominal sensitivity is measured at 23 degrees Celcius. Sensitivity is not a parameter of the measurement capability, only the measured quantity of the measurand.
The dimension of the microphone is the more relevant parameter, corresponding to the geometry of two types of standard laboratory microphones (LSP1 and LSP2).
Microphones have a frequency range, uncertainties are reported for a particular frequency.
Calibration does not depend on reference values, corrections can be applied.
Contact Details
No response
What is the process type?
Measure
Measurand
Pressure sensitivity (ratio of dynamic output voltage to dynamic pressure), typically expressed as a logarithmic level quantity relative to 1 V/Pa.
The measurement capability (primary calibration via reciprocity method ISO 61094-2) is traceable to fundamental physical quantities:
Nominal sensitivity is measured at 23 degrees Celcius. Sensitivity is not a parameter of the measurement capability, only the measured quantity of the measurand.
The dimension of the microphone is the more relevant parameter, corresponding to the geometry of two types of standard laboratory microphones (LSP1 and LSP2).
Microphones have a frequency range, uncertainties are reported for a particular frequency.
Calibration does not depend on reference values, corrections can be applied.
Kind of quantity
Voltage and Pressure
Does the measurand require a special token?
Coefficient
Taxon Name
Measure.Coefficient.Voltage.Pressure.AC.Differential.PressureField
Additional Information
ISO 61094-1 Includes definitions and 61094-2 describes reciprocity method for calibration of standard laboratory microphones.
Parameters
Sensitivity (output quantity)
Geometry (of microphone)
The text was updated successfully, but these errors were encountered: