Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Breath rate adjustment too sensitive #5

Open
kaiem opened this issue Jun 7, 2020 · 2 comments
Open

Breath rate adjustment too sensitive #5

kaiem opened this issue Jun 7, 2020 · 2 comments

Comments

@kaiem
Copy link
Owner

kaiem commented Jun 7, 2020

In the current topology, adjustment of the breath rate is achieved by turning the control knob which changes the rate of vacuum bleed. Reducing the supply vacuum too much in this way increases sensitivity to pilot tube positioning and thereby reduces reliability.

@kaiem
Copy link
Owner Author

kaiem commented Jun 7, 2020

Suggest changing from bleed control to inline throttle in the output tube of the System Control Valve.

@kaiem
Copy link
Owner Author

kaiem commented Jun 7, 2020

Suggest adding a vacuum regulator in the source vacuum line to control the supply vacuum to a known value which would enable the rate adjustment knob position to be consistent and able to be calibrated against breathing rate.